31
Setting the Pulse Mode
Setting the duty
1
2
1
Select the
[Duty]
setting.
2
Press a function key (
F1
to
F5
) or select a button with the mouse.
Up-Down
p.18
Tenkey entry
p.19
Valid duty setting range: 0 to 100 (in 0.1 increments)
Entering a setting of 0 will x the output level to LOW, while entering a setting of 100 will x
the output level to HIGH. The unit will not generate pulse output in either case.
When the duty is set to 100 , the output waveform will be xed at the HIGH level, even when
output is set to OFF (p.33).
Setting a duty value that corresponds to a pulse width that is narrower than 1 s may cause the
pulse to disappear.