72
Pressing
[RUN]
or
[STOP]
may cause the output waveform to exhibit undershoot or overshoot.
When control is set to
[RUN]
or
[PAUSE]
:
U8793: The type of waveform being output cannot be changed. The parameters for the selected
waveform type can be changed. (However, the parameters cannot be changed if any parameter
has been set to “sweep.”)
MR8790: The type of waveform being output as well as associated parameter settings can be
changed.
MR8791: The output mode cannot be changed. Associated parameter settings can be changed.
When
[PAUSE]
is pressed, the unit will continue to output the voltage at the time
[PAUSE]
was
pressed. Subsequently, pressing
[RUN]
will cause output to resume from the phase at the time
that
[PAUSE]
] was pressed.
Output resumes from the phase at
the time that
[PAUSE]
was pressed.
[PAUSE]
is pressed.
[RUN]
is pressed.
While in the
[PAUSE]
state, HIGH-level output will be generated from the external control
terminal’s OUT pin (p.14, p.83).
When output (p.25, p.33, p.57) is set to OFF while in the
[PAUSE]
state:
U8793: Output will be 0 V. When output is turned back on, output will remain 0 V.
M8790: Output will be 0 V. When output is turned back on, output will return to the voltage in effect
while in the
[PAUSE]
state.
MR8791: When operating in pulse mode, output will be LOW-level. When output is turned back
on, output will remain LOW-level.
When in pattern mode, output will be at the same level as the rst pattern. When output is turned
back on, output will remain at the same level as the rst pattern.
When control is set to
[RUN]
or
[PAUSE]
, output can be turned on and off (p.27, p.33, p.39,
p.57), but there may be a disturbance in the waveform when output is turned on or off due to
the output relay’s response.
[RUN]
Output on
Output off
Disturbance occurs in waveform.
After turning output on, no disturbance will occur in the waveform when setting control to
[RUN]
or
[STOP]
.
Control:
[RUN]
Output on
Control:
[STOP]