Editing the Program
Hold setting operation
While the hold function is in effect, HIGH-level output will be generated from the external
control terminal’s OUT pin (p.14, p.83)
When the hold parameter for a step is set to
[On]
, the waveform set by the that step will be
output until the hold is canceled.
When a step for which the hold parameter is set to
[On]
contains a sweep waveform, the
waveform de ned by the end value for each sweep parameter will be output after repeating the
sweep waveform the set number of loops, and that output will continue until the hold state is
canceled.
6ZHHSZDYHIRUPVHWWLQJ
Example
Step 1
Step 2
Step 3
Hold canceled
A LOW-level signal is input to the IN external control terminal.
To cancel the hold state, input a LOW-level signal to the unit’s IN external control terminal.
It takes about 100 s to cancel the hold from the time a LOW-level signal is inputted.
(p.14, p.82)
Switching steps
It takes approximately 20
μ
s to switch to the next step once the previous step ends. The output
voltage during that interval will be the voltage at the time the previous step completed.
The waveform de ned by the sweep end values
will be output while the hold function is in effect.