47
Registering a Waveform
Registering data measured with a Memory HiCorder
1
2
1
Set information about the measurement data to be registered.
Channel
Sets the measurement channel number for the measurement data being
registered (Analog waveform: Ch, Waveform calculation: Z). This parameter
can be set by function key (
F3
,
F4
).
Start Position
Sets the start position for the range of measurement data to be registered.
This parameter can be set by function key (
F1
: entire waveform;
F2
:
waveform between A and B;
F3
: up-down;
F4
: numeric keypad input).
Selecting [
F1
: Entire waveform] will cause the start position to be set to 0,
while selecting [
F2
: Waveform between A and B] will cause the start position
to be set to the position of the A cursor.
Data Count
Sets the number of data points in the measurement data to be registered.
This parameter can be set by function key (
F1
: entire waveform;
F2
:
waveform between A and B;
F3
: up-down;
F4
: numeric keypad input).
Selecting [
F1
: Entire waveform] will cause the data count to be set to the
total number of measured data points, while selecting [
F2
: Waveform
between A and B] will cause the data count to be set to the number of data
points between the A and B cursors (or if only the A cursor is displayed, the
number of data points occurring after the A cursor).
Conv. Ratio
Sets the scaling factor to apply to the voltage values in the measurement
data to be registered when registering the data. The valid setting range for
this parameter is 0.001 to 100.000 . The parameter can be set by function
key (
F3
: up-down;
F4
: numeric keypad input).
Name
Sets the name of the data. Up to 16 single-byte characters or 8 double-byte
characters can be entered.
2
Press (
CH.
SET)
to register the data in the unit’s memory. To cancel the operation
without registering any data, press
(
TRIG.
SET)
or
(
ESC)
.