4.8 Short Circuit Compensation
______________________________________________________________________________________________
4.8.5 When an Error Message Appears and Compensation Has
Stopped
4.8.6 Clearing Compensation Data
When an error message appears and compensation has stopped, short circuit
compensation is turned off.
Check the following points before starting the short circuit compensation
process again:
Check that the test cables are properly connected.
Check that the test cables are properly shorted together with the shorting bar.
(Short circuit compensation cannot be performed while any test sample is
connected to the test cables.)
Check that the test leads are arranged as closely as possible to their
configuration in which measurement will be performed.
During the compensation process, be sure not to disturb the test cables or to
move your hand near them.
From the Short circuit compensation screen, press the
key, and the
display will revert to the Initial screen with the compensation data cleared.