4.15 Setting and Activating the Comparator
______________________________________________________________________________________________
NOTE
The comparator decisions are made in the following order:
1) If the measured value is "OVER FLOW", HI is displayed, and the value is
"UNDER FLOW", LO is displayed.
2) Is the measured value greater than the lower limit value or not?
If not (NG decision result), then LO is displayed.
3) Is the measured value lower than the upper limit value or not?
If not (NG decision result), then HI is displayed.
4) If both 2) and 3) give an affirmative result, then IN is displayed.
No test is performed in order to ensure that the upper limit value is greater
than the lower limit value. Therefore no error message will be displayed
even if you mistakenly interchange the settings for the desired upper limit
value and the desired lower limit value. However, be careful that the
decision process will not operate properly.