177
_____________________________________________________________________________________________
7.5 Test Accuracy
______________________________________________________________________________________________
1 k
Ω
range or more
B x |10 x
Z
x[
Ω
] - Range[
Ω
]|
Basic accuracy =
A
+─────────────
Range[
Ω
]
100
Ω
range or less
B x |Range[
Ω
] -
Z
x[
Ω
]| x 10
Basic accuracy =
A
+─────────────
Range [
Ω
]
Zx: sample of impedance
A, B: basic accuracy coefficient
Accuracy coefficient for Z (%): upper values in the following table
Accuracy coefficient for
θ
(
): lower values in the following table
When the frequency is 1.001 MHz or more, (f [MHz] +3)/4 must be
multiplied to the basic accuracy.
Basic Accuracy Table 2 (0.050 V to 0.100 V)
Содержание 3532-50
Страница 1: ...Instruction Manual For 3532 50 LCR HiTESTER February 2013 Revised edition 10 3532C981 10 13 02H...
Страница 2: ......
Страница 8: ......
Страница 197: ......