167
_____________________________________________________________________________________________
7.3 Time Taken for Testing
______________________________________________________________________________________________
<B>
The time taken for test varies according to the display state
(display ON/OFF):
Display setting
Test time
LCD display and backlight :OFF
-2.3 ms
Voltage and current monitors:OFF
-1.4 ms
LCD display and backlight :ON
Voltage and current monitors:ON
0 ms
(Allowance
1 ms)
<C>
The time taken for calculation varies according as to whether or not both
open circuit compensation and also short circuit compensation are performed:
Open/short circuit compensation
Calculation time
Open and/or short circuit
compensation performed
1.5 ms
Not performed
0 ms
(Allowance
1 ms)
<D>
The time taken for calculation varies according as to whether or not the
comparator is operating:
Comparator operating
Calculation time
Normal testing
0 ms
When the comparator is operating
0.7 ms
(Allowance
1 ms)
<E>
The time taken for calculation varies according as to whether or not the
comparator is operating:
Scaling execution
Calculation time
Normal testing
0 ms
When the scaling is executed
0.8 ms for a parameter
(Allowance
1 ms)