4.15 Setting and Activating the Comparator
______________________________________________________________________________________________
Numeric
keypad
OFF key
Units are
percent (%)
NOTE
Percentage set value
= Reference value + |Reference value|
x
________________
100
Upper-limit
comparison value
3.
Select the units using one of the units keys:
:
step the units up
:
step the units down
p -- n --
μ
-- m -- (none) -- k -- M
<= Units down
Units up =>
Pressing the
key causes the upper limit value to be set and the screen
display to return to the Conditions setting screen.
4.
When the
key is pressed, a screen for setting the upper limit value
will appear. Use the numeric keypad to input the upper limit value as a
percentage relative to the reference value.
In order to set no upper limit value, press the
key.
The setting can be made in the range : -999.99% to +999.99%
When the setting has been established as above, press the
key to return
to the Conditions setting screen with the upper limit value set.
The actual internal operation consists of calculating the upper-limit value of
comparison using the equation given below, and comparing it to the
measurement value to enable a decision to be made.
| |: Denotes an absolute value