23
_____________________________________________________________________________________________
3.3 Basic Measurement
______________________________________________________________________________________________
3.3.5 Setting Open Circuit Compensation
Before Setting
After Setting
Initial Screen
Menu Screen
The following explanation shows, as an example, how to set up ALL open
circuit compensation, with which open circuit compensation values for all of
the test frequencies are determined and adjusted.
Setting procedure
1.
Arrange the test leads as closely as possible to
their configuration in which measurement will
be performed, and make sure that the HIGH
and LOW leads are not contacted together.
For details, refer to Section 4.7.2, "Selecting
the Compensation Method".
2.
On the Initial screen, press the
key, and
the Menu screen will be displayed.
3.
Press the
key on the Menu screen, and
the Open circuit compensation screen will be
displayed.