Cleaning, disinfection and maintenance
Cleaning
8
Cleaning, disinfection and maintenance
DANGER
Danger due to insufficient cleaning!
In the case of insufficient cleaning or non-observance of the
cleaning regulations, there is a risk of contamination for the
user.
−
Observe the specified cleaning regulations.
−
Observe the laboratory rules (TRBAs, IfSG, hygiene plan,
etc.) for handling biological agents.
−
Wear personal protective equipment when cleaning the
device (gloves, protective clothing, etc.).
Observe the following during cleaning and disinfection:
■
The device and the rotor may not be cleaned in dishwashers.
■
The device and the rotor may not be cleaned under running water, in
water baths or with compressed air.
■
The device and the rotor may not be sterilized in autoclaves, since this
can destroy the electronic components.
■
Only manual cleaning may be performed.
■
To avoid signs of corrosion due to cleaning agents and disinfectants, it is
imperative that the special instructions for use from the manufacturer of
the cleaning agent and disinfectant be observed.
DANGER
Danger due to insufficient maintenance or maintenance not
performed in a timely manner!
There is a risk of collision inside the device due to insufficient
maintenance or maintenance not performed in a timely
manner.
−
Have maintenance done at the specified intervals.
−
Check the rotor for a firm fit before using the device.
Correctly tighten the fastening screw. If the rotor has
play, it must be replaced.
−
Check the device for visible damage or deficiencies
before use. If you discover deficiencies or damage, put
the device out of operation and immediately inform an
authorized service technician.
8.1
Cleaning
Personnel:
■
Authorized skilled personnel with special instruction