Commissioning
5
Commissioning
■
Set up and level the device at a suitable location so that it is stable.
During setup, the required safety zone of 300 mm around the device
must be complied with in acc. with IEC 61010-2-020 Ed.2 2006 "Safety
requirements for electrical equipment for measurement, control, and lab-
oratory use – Part 2-020: Particular requirements for laboratory centri-
fuges.
WARNING
During a centrifugation run, no personnel, hazardous sub-
stances or objects may be within a safety zone of 300 mm
around the device in acc. with IEC 61010-2-020 Ed.2 2006
"Safety requirements for electrical equipment for measure-
ment, control, and laboratory use – Part 2-020: Particular
requirements for laboratory centrifuges.
■
Ventilation openings may not be blocked. A distance of 300 mm to the
air slots and ventilation openings must be complied with.
■
Before commissioning, check to make sure the mains voltage is correct
according to the nameplate and that the mains cable is connected prop-
erly.
■
Only trained and instructed personnel may commission and use the
device.
Every time before the device is used, make sure that
■
the device is dry (no moist residues from cleaning agents, etc.).
■
no objects are lying in the centrifuge chamber under the rotor.
■
the rotor position corresponds to the home position.
■
the accessories are correctly connected to each other (leak-proof) and
the rotor is correctly inserted.
■
the shown displays are correct.
5.1
Unpacking the device
Personnel:
■
Authorized skilled personnel with special instruction
The device is packaged in a box.
When lifting up the device, note the specified weight and
only lift the device with a suitable number of helpers to
avoid injuries.
WARNING
Warning - Position changed due to vibration!
The device might fall down if the set-up surface vibrates or if
it is not flat or stable enough.
−
Only place the device on a stable, flat surface.
−
The set-up surface must reliably bear the weight of the
device.