41 / 88
en
Rev. 03 / 02.2021
AB3200
Operation
Fig. 22: Rotor cavity
symmetrically arranged
7.
Place the adapter (in the case of rotor for dual centrifugation) into the
rotor cavities.
8.
Put the device into operation.
6.10 Cooling
The temperature can be set from -20 °C to +40 °C / -4 °F
to +104 °F. The device displays the temperature of the cen-
trifuge chamber. The actual temperature of the sample can
vastly differ from the displayed temperature.
When the rotor is at a standstill, with the lid closed, the centrifuge chamber
is cooled to the preselected temperature if this is lower than 20°C / 68°F.
During standby cooling, the preselected temperature is displayed.
Dual centrifugation should fundamentally be applied with
cooling in order to dissipate the heat that arises during the
process and to prevent the samples from overheating.