73 / 88
en
Rev. 03 / 02.2021
AB3200
Troubleshooting
Fault description
Cause
Remedy
CONTROL-ERROR 22, 25.1–
25.4; SER I/O - ERROR 31,
34, 36; ° C * - ERROR 51, 53
– 55, 97, 98; FC INIT ERROR;
FC VERSION ERROR; FATAL
EEPROM ERROR 1-5
Error, defective electronics.
Perform a MAINS-RESET
Chapter 9.1 ‘Performing a mains
reset’ on page 74. If the error persists
after a mains reset, contact a service
technician.
° C * - ERROR 52.0, 52.1
Overtemperature in the centrifuge
chamber. Overtemperature switch
defective. Temperature sensor
defective.
Perform a MAINS-RESET
Chapter 9.1 ‘Performing a mains
reset’ on page 74. If the error persists
after a mains reset, contact a service
technician.
FU / CCI - ERROR 60, 61.2-
61.20, 61.128 - 61.131, 62;
SENSOR-ERROR 90
Error, defective electronics, motor. Perform a MAINS-RESET
Chapter 9.1 ‘Performing a mains
reset’ on page 74. If the error persists
after a mains reset, contact a service
technician.
SENSOR-ERROR 91 - 93
Error, defective unbalance sensor.
Perform a MAINS-RESET
Chapter 9.1 ‘Performing a mains
reset’ on page 74. If the error persists
after a mains reset, contact a service
technician.
NO ROTOR OR ROTORCODE
ERROR
No rotor installed.
Install rotor
and installing the rotor’ on page 39.
Tachometer defective.
Perform a MAINS-RESET
Chapter 9.1 ‘Performing a mains
reset’ on page 74. If the error persists
after a mains reset, contact a service
technician.
FU / CCI - ERROR 61.1
Mains voltage too low. Error,
defective electronics, motor.
Check the mains voltage.
Perform a MAINS-
RESET
mains reset’ on page 74. If the error
persists after a mains reset, contact a
service technician.
MAINS INTERRUPT
Mains interruption during the cen-
trifugation run. (The centrifugation
run was not finished.)
WRONG ROTOR !!!
Rotor is not approved for the
device.
Install approved rotor
‘Removing and installing the rotor’
N > ROTOR MAX
Speed in the activated program
higher than the maximum rotor
speed.
Check the speed and correct.