Troubleshooting
Fault description
Cause
Remedy
N > ROTOR MAX
The rotor was changed. The
installed rotor has a higher max-
imum speed than the previously
used rotor, and it has not yet
been detected by the rotor detec-
tion function.
Set a speed (up to the maximum
speed of the previously used rotor).
Press the
[START] key to detect the
rotor.
N > ROTOR MAX in Prog
There is a program in the dis-
played program location whose
speed is higher than the max-
imum speed of the rotor.
Check the speed and correct.
The rotor was changed. The
installed rotor has a higher max-
imum speed than the previously
used rotor, and it has not yet
been detected by the rotor detec-
tion function.
Set a speed (up to the maximum
speed of the previously used rotor.
Press the
[START] key to detect the
rotor.
Runtime 00:00 in Prog
There is a continuous operation
program at the displayed program
location.
Replace the continuous operation pro-
gram in the program link with a pro-
gram with a time preselection.
Empty Program
There is no program linkage
saved on the displayed program
location.
Call up a program linkage.
Ramp Unit Time in Prog
A program with start-up and/or
run-down time is located at the
displayed program place.
Replace the program in the program
link with a program with start-up and
braking stage.
Acc time > Run time
The set start-up time is longer
than the run time.
Set a start-up time which is shorter
than the run time.
9.1
Performing a mains reset
1.
Switch off the mains switch (switch setting
‘0’).
2.
Wait for at least 10 seconds and then switch the mains switch back
on (switch setting
‘I’).