37/65
Carrying out a test run
↓
START
PULSE
•
Press the key.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
STOP
OPEN
•
Press the key.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
•
Load a rotor place with the switch-off weight (for the
switch-off weight, see the table on page 33).
For angle rotors, use a centrifuging container. In
the case of hematocrit rotors, place the switch-
off weight in the padded shell of the rotor place.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
•
Close the lid.
t/min:s
RPM
>RCF<
↓
•
Press the key.
START
PULSE
A short centrifugation run is carried
out to check the imbalance switch-
off.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
If no imbalance is detected during
the test run, the following error
message will appear.
Press any key to clear the error
message.
Possible cause:
•
The imbalance switch-off is
incorrectly set.
•
The switch-off weight used for the
test run is not heavy enough.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
•
Open the lid.
t/min:s
RPM
>RCF<
↓
STOP
OPEN
•
Press the key once to exit the
"
Imbalance" menu or twice to exit the
" SELECT MENU ".
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<