36/65
START
PULSE
•
Only press the key if a PWM offset value
is displayed.
Example:
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
•
Load a rotor place with the setting weight (for the
setting weight, see the table on page 33).
For angle rotors, use a centrifuging container.
In the case of hematocrit rotors, place the
setting weight in the padded shell of the rotor
place.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
•
Close the lid.
t/min:s
RPM
>RCF<
↓
•
Press the key.
START
PULSE
Two short centrifugation runs are
carried out to set the imbalance
switch-off.
Afterwards, a test run can be carried
out to check the imbalance switch-
off.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
Ending the imbalance adjustment
↓
STOP
OPEN
•
Only press the key if no test run is to be
carried out.
The imbalance adjustment will be ended.
t/min:s
RPM
>RCF<
↓
Continued on next page