29/65
7.2.8
Resetting the cycle counter to zero
After the rotor has been exchanged, the cycle counter must be reset to zero again.
The cycle counter may only be reset to zero if the rotor has been exchanged for a new rotor first.
With the rotor at a standstill, the cycle counter can be reset as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
SELECT
•
Press and hold the button for eight
seconds.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key until the following is
displayed.
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
Example:
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key until the following is
displayed:
Cyc sum :
Number of completed run
cycles.
t/min:s
RPM
>RCF<
↓
RCF
•
Press the key.
t/min:s
RPM
>RCF<
↓
t
•
Press the key.
The number of completed run cycles is
reset to zero.
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the button to save the setting.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
↓
STOP
OPEN
•
Press the key twice to exit the "
Time &
Cycles"
menu or press three times to exit the
" MACHINE MENU ".
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<