43/65
7.4.6
Deleting the centrifugation parameters in the EEPROM
The centrifugation parameters in the EEPROM are reset to their default values.
Centrifugation parameters
Default value
Runtime t/min,
t/sec
1:00
Speed RPM
1000
Relative centrifugal force
RCF 56
Centrifuging radius
RAD/mm
50
Brake stage
DEC fast
With the rotor at a standstill, the centrifugation parameters can be deleted as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
START
PULSE
+
STOP
OPEN
+
•
Keep both keys pressed and
switch on the mains switch at
the same time.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key until the following is
displayed:
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
t/min:s
RPM
>RCF<
↓
t
•
Set
yes
with the keys.
yes :
Delete centrifugation data
no :
Do not delete centrifugation data
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
Repeat confirmation
t/min:s
RPM
>RCF<
↓
t
•
Set
yes
with the keys.
yes :
Delete centrifugation data
no :
Do not delete centrifugation data
t/min:s
RPM
>RCF<
↓
Continued on next page