18/65
FC ERROR 61.64,
FC ERROR 61.128,
FC ERROR 61.129,
FC ERROR 61.131,
FC ERROR 61.132,
FC ERROR 61.140,
FC ERROR 61.141,
FC ERROR 61.142
•
Error description:
Internal error of the frequency converter.
•
Clear error message:
Carry out a MAINS RESET.
•
Cause of error:
Electronics (A1) is defective.
IMBALANCE
•
Error description:
An impermissible imbalance was detected. Braked run-down of the rotor until it comes to a standstill.
•
Clear error message:
Wait until the rotor has come to a standstill and the symbol (lid locked) is displayed. Afterwards, unlock the lid
or carry out a MAINS RESET.
•
Cause of error:
–
The rotor is unevenly loaded.
–
The three rubber/metal bearings on the motor are damaged. Check the three rubber/metal bearings for wear
or cracks.
–
The imbalance switch-off is maladjusted. Adjust the imbalance switch-off; see chapter 7.4.2.
–
Electronics (A1) is defective.
FC ERROR 61.153
•
Error description:
The imbalance system is not in working order.
•
Clear error message:
Carry out a MAINS RESET.
•
Cause of error:
–
There is no holder with magnet fastened to the motor for detecting imbalance.
–
The magnet for detecting imbalance is defective or is no longer in the holder.
–
The holder is placed such that the magnet is no longer above the hall sensor.
–
Electronics (A1) is defective.
–
The three rubber/metal bearings on the motor are so damaged that the magnet is no longer above the hall
sensor.