21/65
7.2.1
Querying system information
The following system information can be queried:
•
Centrifuge model
•
Centrifuge program version
•
Frequency converter type
•
Frequency converter program version
With the rotor at a standstill, the query can proceed as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key.
SELECT
•
Press and hold the button for eight
seconds.
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
t/min:s
RPM
>RCF<
↓
START
PULSE
•
Press the key.
Centrifuge model.
Example:
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
Centrifuge program version.
Example:
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
Frequency converter type.
Example:
t/min:s
RPM
>RCF<
↓
SELECT
•
Press the key.
Frequency converter program version.
Example:
t/min:s
RPM
>RCF<
↓
STOP
OPEN
•
Press the key twice to exit the "
Info"
menu or press three times to exit the
" MACHINE MENU ".
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<