4.3
Motor (M1) / Speed sensor (B1)
•
The motor is a 2-pole three-phase asynchronous motor (1 pair of poles).
•
The motor is protected against overheating by an overtemperature switch. The overtemperature switch is
integrated in the motor and cannot be replaced.
•
The electronics (A1) evaluates the overtemperature switch.
•
The speed sensor (B1, speedometer) is screwed to the bottom of the motor. The speed signal (1 pulse per
revolution) will be triggered by a magnet located in the motor axle.
•
The electronics (A1) monitors and controls the speed of the motor.
•
In addition the speed of the motor is also monitored by the electronics (A2).
•
The holder of the magnet for the imbalance detection is attached to the motor.
4.4
Motor-driven lid lock (A4)
•
The lid can be opened only if the electronics (A1) has detected rotor standstill.
•
After pressing the key
STOP
OPEN
the lid opens motor-driven.
•
The centrifugation run can only be started when the lid is closed and the lid lock has locked. A microswitch on
the lid lock detects the position of the lid lock (open/closed) and report it to the electronics (A1).
4.5 Imbalance
detection
•
An hall sensor monitors the imbalance. This hall sensor is located on the electronics (A1).
•
The magnet for the imbalance detection is located in a holder which is attached to the motor. The holder is
placed in such a way that the magnet is located directly above the hall sensor.
•
Imbalance is detected only in running mode (run up, centrifuging and braking).
•
If impermissible imbalance is detected, the motor switches off and the rotor slows down braked until it stops.
5 Troubleshooting
procedures
•
Fuses in installation in which centrifuge is installed are intact.
•
Mains input fuses of centrifuge are intact.
•
Mains voltage present at (see connecting diagram):
EBA 200:
−
Connecting cable
−
Appliance plug/mains switch
−
Electronics (A1), plug S102L and S102N
EBA 200 S, HAEMATOKRIT 200:
−
Connecting cable
−
Appliance plug/mains switch on the electronics (A3), plug SL1 and SN1
−
Electronics (A1), plug S100, pin 1 (L1) and pin 3 (N)
•
Look for the displayed error message in chapter 6 on page 8.
•
Remedy the error according to the instructions.
•
Carry out a functional check after every repair and whenever a component is replaced, see pg. 48, chapter 8.