ERS COMPACT
Group
Parameters & description
Timeouts
TCP keep-alive (slot) / TCP keep-alive (internal)
Activates/deactivates keep-alive network functionality for the internal Ethernet port and
for an a standard Ethernet module (IF-AB-ETHxx), if installed in the slot. The “keep-alive”
network packets are used to keep the socket connection open. As long as keep-alive
is valid in the network, the device will disable the Ethernet timeout. Also see below at
Timeout ETH
.
Timeout USB/RS232
Defines the max. time between two subsequent bytes or blocks of a transferred message.
For more information about the timeout refer to the external programming documentation
“Programming ModBus & SCPI”. Default value:
5 ms
, Range: 5...65535
Timeout ETH
Defines a timeout after which the device would close the socket connection if there was
no command communication between the controlling unit (PC, PLC etc.) and the device
for the adjusted time. The timeout is ineffective as long as option “TCP keep-alive” is en
-
abled and the keep-alive network service is running. Default value:
5 s
, Range: 5...65535
Interface monitoring / Timeout Interface monitoring
Activates/deactivates the interface monitoring (see section
“3.6.3.5. Interface monitoring”
).
Default values: off,
5 s
/ Range: 5..65535
Protocols
Communication protocols
Enables or disables SCPI or ModBus communication protocols for the device. The change
is immediately effective. Only one of both can be disabled.
ModBus specification compliance
Allows to switch from
Limited
(default setting) to
Full
which makes the device send mes-
sages in ModBus RTU or ModBus TCP format which fully comply to the specification and
are compatible to softwares available on the market. With
Limited
the device would still
use the old, partially wrong message format (see programming guide for details).
3.5.3.7
Menu “HMI Setup”
These settings refer exclusively to the control panel (HMI).
Group
Parameters & description
Language
Selection of the display language between (default: English)
Sound
Key sound
Activates or deactivates sounds when touching a touch area in the display. It can acous-
tically signal that the action has been accepted.
Alarm sound
Activates or deactivates the additional acoustic signal of an alarm or user defined event
which has been set to “Action = Alarm”. See also
“3.7. Alarms and monitoring”.
Clock
Internal clock and date setup
Backlight
Backlight off after 60s
The choice here is whether the backlight remains permanently on (default) or if it should go
off when no input via screen or rotary knob is done for 60 s. As soon as there is input, the
backlight returns automatically. Furthermore, the backlight intensity can be adjusted here.
Lock
See
““3.8. Locking the control panel (HMI)”
and
“3.9. Locking the “Limits” and “Profiles””