Chapter 2 – Installation
Hardy Process Solutions HI 6200 Weigh Processor User Guide
17
2.4 Climatization
An HI 6200 shipped or stored at extreme temperatures requires time to adjust to operating
temperatures before startup.
If the HI 6200 arrives in hot or cold weather, do not unpack it until it has been allowed to
reach room temperature (one to two hours).
Immediately exposing the HI 6200 to warm temperature can cause condensation to occur,
which could damage the electronics.
If you notice any condensation, allow the HI 6200 to stand unattended for one to two hours,
and then unpack it.
2.5 Handling Precautions
Observe the following precautions when handling the HI 6200:
Wear an approved wrist-strap grounding device when handling the instrument.
Touch a grounded object or surface to rid yourself of any electrostatic discharged prior to
handling the instrument.
Handle the instrument from the bezel in front away from the connector. Never touch the
connector pins.
Do not install the instrument right next to an AC or high voltage DC module.
Route all the load voltage cables away from high voltage cables.