Chapter 4 – Ways to Set Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
28
4.2 Using the Embedded Webserver
The embedded Webserver is a graphical point-and-click interface that can be accessed
using a computer that has an Internet connection. Using the embedded Webserver, you can
access the same instrument settings and parameters available from the front panel touch
screen plus the following additional features:
Advanced diagnostics
Parameter cloning
Firmware upgrade
To set up the HI 6200 using its embedded Webserver:
1. If the HI 6200 is powered on, power it off.
2. Connect either end of a straight-through or crossover Ethernet cable to the RJ-45
connector on the top of the HI 6200. Connect the other end of the cable to a Microsoft
Windows computer.
3. Power up the HI 6200. The instrument automatically negotiates settings suitable for a
variety of operating systems and network configurations.
4. Wait approximately 15 seconds, and then check whether the green light of the Ethernet
port is blinking. If it is blinking, the computer can communicate with the HI 6200 and you
can skip to
step 5
. Otherwise, perform the following steps:
a. The HI 6200’s default static IP address is 192.168.0.100. To communicate with the
instrument, configure the computer’s IP address so the network identifier (first three
octets in the IP address) are the same as the HI 6200 and the node identifier (last
octet) is a number from 0 to 255 other than 100. For example: 192.168.0.101.