Chapter 4 – Ways to Set Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
40
Status Error code 4
Motion
Status code FF
cmd in progress
TARE CMD
Hex value: 0x02
Decimal value: 2
Performing a tare command changes the net weight to “0” and moves the tared value into
the “tare weight” parameter. Write the hexadecimal value 0x02 to the command register to
zero the net weight. If this command succeeds, the Cmd Status register (lower 8 bits of
CmdStatusNCount
) reads 0.
Status Error code 1
Fail
Status Error code 2
ADC Failure
Status Error code 4
Motion
Status code FF
cmd in progress
CAL LOW CMD
Hex value: 0x64
Decimal value: 100
Write the hexadecimal value 0x64 to the command register to perform the low step of a
traditional calibration. If this command succeeds, the status register reads 0.
Status Error code 1
Fail
Status Error code 2
ADC Failure
Status Error code 4
Motion
Status code FF
cmd in progress
CAL HIGH CMD
Hex value: 0x65
Decimal value: 101
Write the hexadecimal value 0x65 to the command register to perform the high step of a
traditional calibration.
Status Error code 1
Fail
Status Error code 2
ADC Failure
Status Error code 4
Motion