Chapter 4 – Ways to Set Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
51
A zero indicating the command passed.
A value of 0xFF indicating the command is in process.
An error code indicating the reason the command failed (see Table 4-5 on page 43).
The upper 8 bits of the register are a cyclic “measurement update count,” which increment
by one each time a new measurement value is taken, following a 0 to 255 then repeat cycle.
If this value remains the same in two consecutive reads from the instrument, the
communication or the measurement function has failed and appropriate action must be
taken.
Parameter ID
is an echo of the value sent in the output table.
Parameter Value
is the
value for the specified Parameter ID.
Instrument Status
is a 16-bit value that provides the current state of all the major
functions within the instrument. They reflect the status of all the major functions and should
be used with the “measurement update count” to determine the health of the instrument.
The
Net Weight
and
Gross
Weight
values are always provided.
4.3.6.3 Default Parameters
Table 4-8 shows the default parameters.
Table 4-8. Default Parameters
Configuration Table
Data Type
Defaults
Range
Gravity Correction
FLOAT
1.0
.9-1.2
Motion Tolerance
FLOAT
10.0
.00001-999999.99
Zero Tolerance
FLOAT
10.0
.00001-999999.99
Tare Weight
FLOAT
0.0
Reference Weight
FLOAT
0.0
.00001-999999.99
Span Weight
FLOAT
1000.0
.00001-999999.99
AutoZero Tolerance
FLOAT
10.0
0 -999999.99
Enable AutoZero Tracking
BYTE
0
0-1
0 = off
1 = on
Unit
BYTE
1
0-5
0 = oz
1 = lb
2 = ton
3 = g
4 = kg
5 = t
DRAFT