Chapter 4 – Ways to Set Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
33
4.2.2 Restoring from “KnownGood” Parameters
After you use the procedure in section 4.2.1 to save a last-known good configuration, you
can restore the HI 6200 to that configuration if necessary.
1. Open a web browser on the computer, enter the static IP address of the HI 6200
instrument (192.168.000.100), and press the Enter key. The Home page appears (see
Figure 4-1 on page 30).
2. Click
Configuration
, and then click
Diagnostics
.
3. Click
Parameters
.
4. Click
Known Good Cfg
.
DRAFT