Chapter 4 – Ways to Set Scale Parameters
Hardy Process Solutions HI 6200 Weigh Processor User Guide
43
Set Default Parameters
Hex value: 0x94
Decimal value: 148
Write the hexadecimal value 0x94 to the command register to set all parameters and
calibration back to default settings.
Status Error code 1
Fail
Table 4-4. Instrument Status Word Bits
Bit
Description
0
A/D converter error - bad input from the load sensor.
1
A/D converter failure - no output from the converter to the processor.
2
Motion - indicates weight is in motion (changing).
3
Center of Zero
6
Calibration in Progress
7
Error parameter ID Not Found
Table 4-5. Command Status Return Value
Return Value
Description
0
Success
1
Fail
2
Fail - ADC error and ADC failure
3
Fail - out of tolerance
4
Fail - motion
5
Fail - no C2 load cells found
6
Fail - C2 capacities not equal
7
Fail - non-Hardy C2 load sensor
8
Fail - not enough counts between Cal low and Cal high weights
11
Fail – param value too high
12
Fail – param value too low
13
Fail – not allowed
128
Fail – Parameter ID not found