Hardy Process Solutions HI 6200 Weigh Processor User Guide
24
3.1 Initial Setup
The following steps describe the minimum configuration procedures to perform when setting
up the HI 6200 for the first time. Perform these procedures in the order shown.
For detailed descriptions of scale parameters and how to use them when setting up a scale
system, see Chapter 5.
1. Choose a calibration method:
C2
or
Hard Cal.
Recommended Method: Perform C2 – Electronic Calibration:
a. Remove all weight from the scale.
b. Press
Do C2 Cal
.
c. Place a verification weight on the scale to ensure the weight reading matches and that
calibration is successful.
Alternate Method: Perform Hard Cal
-
Traditional
a. Input the span weight (about 80% of the scale capacity).
b. Input the Cal Low Weight with nothing on the scale. Confirm that Cal Low Weight is 0.
c. Press
Do Cal Low
to set the low point on the calibration curve.
d. Place the span weight onto the scale.
e. Press
Do Cal Hi
.