P/N 53063-30, Manual revision: B
2-13
Describing the Cell Saver 5+ System Components
Effluent line sensor
The device is equipped with an effluent line sensor which monitors the
quality of the bowl effluent (fluid to waste). The line sensor groove contains
two optical sensors which are used to:
!
Provide information regarding the wash status and advance the sys-
tem to the EMPTY state once the proper effluent quality has been
reached after a “minimum” wash volume has been introduced.
!
Regulate the pump speed in the FILL and WASH states to ensure opti-
mal efficiency with the cell saving process.
Waste bag weigher
The waste bag weigher determines the weight of the fluid in the waste bag
and alerts the operator first when it is almost full, then when it is full.
In the AUTOMATIC mode, the message “
Please empty waste bag
” is
displayed when approximately 8 liters of fluid are in the bag for the Latham
bowl sets and approximately 4.5 liters for the 70mL Bowl set. In both the
AUTOMATIC and MANUAL modes, the following message will be
displayed when 9 liters or more of fluid are in the waste bag for the Latham
bowl sets and 4.5 liters or more for the 70mL Bowl set.
Reservoir level sensor
The device is equipped with a reservoir level sensor which monitors the
weight of the reservoir throughout the procedure to estimate the volume of
fluid it contains. This sensor is designed to trip the system into the FILL mode
when an appropriate amount of solution has been aspirated from the oper-
ating field. The preset trip point will depend on the size of the bowl. The
operator can modify the trip points to adjust to varying situations (example:
800 ml for a standard bowl set LN263).
@-&0%+A-6+B388
28%-&%+%./0:+,-&0%+5-6