P/N 53063-30, Manual revision: B
4-13
Preparing for a Cell Saver 5+ Procedure
For the 70mL Bowl set
"
Locate the line sensor tubing section on the effluent tubing.
"
Thread the line sensor section through the effluent line sensor groove.
"
Ensure that the tubing is deeply inserted as depicted.
Figure 4-12, Installing the line sensor tubing: 70mL Bowl set
For all sets
"
Close the centrifuge cover.
Hanging the bags
"
Hang the reinfusion bag on the IV pole as depicted.
Figure 4-13, Hanging the reinfusion bag
"
Close the two small clamps on the each reinfusion line to the patient.
"
Verify that the bag is securely connected to the blue-coded line.
"
Ensure that the large clamp on the blue-coded line is open.