P/N 53063-30, Manual revision: B
5-19
Cell Saving using Automatic Operation
V
IEWING
THE
C
ELL
S
AVER
5+
SYSTEM
STATUS
The following section provides information about the how procedure statis-
tics are presented on the CS5+ display screen.
Figure 5-13, Example of a FILL state screen display
The upper left section displays the current operating status of the device. In
this example, the CS5+ device is in the FILL state and fluid is being pumped
from the reservoir into the bowl at 600 ml/min.
The right section provides the current procedure statistics for this example as
follows:
!
The CS5+ device is under automatic control.
!
Two bowls have been processed.
!
2000 ml of saline solution has been used during the wash cycles.
!
A total volume of 2525 ml of fluid has been processed from the reser-
voir.
!
400 ml of packed RBCs have been sent to the reinfusion bag.
The operator can make minor adjustments to increase or decrease pump
speed without reprogramming the system by using the slew keys of the
pump control keypad section. These temporary changes are valid only for
the present operating state (e.g.FILL). As the system enters the next state, the
programmed values are reinstated. The temporary pump speeds may range
from 0 to 1000 ml/min adjusted by 25 ml/min increments.
The Pause key can be pressed to stop the pump.
E