P/N 53063-30, Manual revision: B
5-21
Cell Saving using Automatic Operation
Note: The CS5+ device can usually process 500 ml of fluid in approximately
one minute.
!
The product volume displayed to the operator will list approximately
20 ml more than what is contained in the reservoir because 20 ml of
fluid remain trapped in the blue line until the end of the procedure.
The operator should remember to subtract this 20 ml volume from the
display readout until the blue line is drained at the end of the proce-
dure and the 20 ml are recovered.
For example, if 400 ml of product volume is listed on the display
screen, there will actually be 380 ml in the reinfusion bag until the 20
ml are emptied from the blue line at the end of the procedure.
Warning: Accurate volume accounting and WASH monitoring functions
require that all slide clamps be open and that no kinks, twists or flat spots
are present in the tubing.