
P/N 53063-30, Manual revision: B
6-10
Cell Saving using Manual Operation
S
UMMARIZING
PARAMETER
VARIATIONS
The following table summarize the different values for default settings used
throughout the various states of the cell saving protocol and which are
recommended for use during manual operation.
Table 6-1, Pump rate range
State
70mL Bowl
Latham 125
Latham 225
FILL
125 ml/min
300 ml/min
600 ml/min
WASH
100 ml/min
300 ml/min
500 ml/min
EMPTY
100 ml/min
100 ml/min
300 ml/min
Table 6-2, Wash volume per cycle
Procedure
type
70mL Bowl
Latham 125
Latham 225
Standard
300 ml
750 ml
1000 ml
Orthopedic
400 ml
1000 ml
1500 ml
Table 6-3, Maximum centrifuge speed
State
70mL Bowl
Latham 125
Latham 225
FILL
7000 rpm
5650 rpm
5650 rpm
WASH
7000 rpm
5650 rpm
5650 rpm