P/N 53063-30, Manual revision: B
9-2
Maintaining the Cell Saver 5+ Device
P
ROVIDING
AN
OVERVIEW
OF
NORMAL
MAINTENANCE
The Cell Saver 5+ device is easy to use and has been designed to require
minimal maintenance. Normal maintenance will consist of:
!
Cleaning the system and any spilled blood.
!
Ensuring that the pump rollers are clean and free-rolling.
A record should be kept regularly of the date and type of maintenance
performed.
Haemonetics recommends a maintenance visit once or twice a year by an
authorized Haemonetics technician who will perform a series of mainte-
nance controls and fine-tune the device for maximum performance. The
local Haemonetics representative can be contacted for details of a Preven-
tive Maintenance Plan (PMP) for the CS5+ equipment.
Warning: To eliminate the potential danger of electrical shock, the CS5+
device must be disconnected from any power supply prior to servicing or
cleaning the system.
The following list described the basic material required for routine cleaning
and maintenance:
!
Disinfectant cleaning solution compatible for use with the CS5+
material, specific for blood-born pathogens.
!
Warm water.
!
70% Isopropyl alcohol.
!
Lint-free gauze or cloth (for cleaning and drying).
!
Cotton swabs.
!
Protective gloves.
!
Philips-head screw driver.
If there is no established institutional policy for decontamination, Haemon-
etics recommends that blood spills be cleaned with a 10% bleach solution
applied with lint-free gauze.
If there is any question about the compatibility of a cleaning solution with
the CS5+ material, the operator should contact Haemonetics.