P/N 53063-30, Manual revision: B
A-14
Providing Reference Information
IEC 60601-1-2:2001, Table 202: Guidance and manufacturer’s declaration - electromagnetic
immunity
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast tran-
sient burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differen-
tial mode
± 2 kV common
mode
± 1 kV differen-
tial mode
± 2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, volt-
age variations on
power supply input
lines
IEC 61000-4-11
< 5%
U
T
(> 95%
dip in
U
T
) for 0.5
cycles
40%
U
T
(60%
dip in
U
T
)
for 5 cycles
70%
U
T
(30%
dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95%
dip in
U
T
) for 5
sec
< 5%
U
T
(> 95%
dip in
U
T
) for 0.5
cycles
40%
U
T
(60%
dip in
U
T
)
for 5 cycles
70%
U
T
(30%
dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95%
dip in
U
T
) for 5
sec
Mains power quality should be
that of a typical commercial or
hospital environment.
If the operator of the CS5+ sys-
tem requires continued opera-
tion during power mains
interruptions, it is recom-
mended that the CS5+ system be
powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.