P/N 53063-30, Manual revision: B
2-9
Describing the Cell Saver 5+ System Components
Pump control keys
Figure 2-8, Pump control key section
These keys can be used to regulate the pump speed during any CS5+ proce-
dure, however the increments of change will vary between the Cell Saving
and Sequestering protocols, as described in the following table:
Key
Purpose
!
Cell Saver:
Used to increase the pump speed by incre-
ments of 25 ml/min up to a maximum of 1000 ml/min.
Sequester:
Used to increase the pump speed.
Fill:
increments of 10 ml/min up to a maximum of 250.
Empty, Conc:
increments of 25 ml/min to a max of 1000.
"
Cell Saver:
Used to decrease the pump speed by incre-
ments of 25 ml/min to a minimum of 0.
Sequester:
Used to decrease the pump speed.
Fill:
increments of 10 ml/min to a minimum of 0.
Empty, Conc:
increments of 25 ml/min to a minimum of
0.
Pause
Cell Saver / Sequester:
Used to stop only the pump (not
the centrifuge).