P/N 53063-30, Manual revision: B
7-6
Sequestering using the Cell Saver 5+ System
P
ROCESSING
FROM
BLOOD
BAGS
Using this method to sequester plasma, the operator will first need to collect
whole blood from the patient via a short intravenous or arterial cannula into
blood bag(s) containing anticoagulant solution.
Warning: Only one (1) unit of blood should be collected and processed at a
time. Unless directed by a physician, the second unit of blood should not
be drawn from the patient until the first unit of packed autologous red
blood cells has been returned.
Then operator should continue as follows:
"
Spike the blood collection bag with the yellow-coded wash line bag
spike on the standard CS5+ disposable set.
"
Clamp the unused wash line spike section.
Filling the bowl
The operator can now begin to draw blood into the bowl from the blood
bag.
"
Press the Start key to initiate filling the bowl.
The pump will start drawing blood at 60 ml/min. As the bowl fills, plasma
will be visible as the first layer followed by a white band (platelets), then red
blood cells. The following information will be displayed:
Figure 7-5, Example of the FILL state screen display
The operator should observe the centrifuge bowl as it fills and allow air to
flow from the circuit to the waste bag.
When plasma reaches the effluent line:
"
Unclamp the clear line to the plasma collection bag.
"
Clamp the line to the waste bag.