77
Communication Menu
10.1.1 Data available
All individual data are separated by at least one tabulation character (ASCII code=0x09).
For the cyclic measurements, the data format is detailed. For the files, only one example for
each file is given to explain the data format.
Cyclic measurements
1.
If the option "Mode = standard" is chosen, the following message is sent:
with:
\t ................................................the ASCII tab character: code=0x09
\r ................................................the ASCII Carriage Return character: code=0x0D
\n ..............................................the ASCII Line Feed character: code=0x0A
CHn ..........................................the 2 ASCII characters "CH" + the channel no.
Gas ............................................the gas concentration.
Gas Unit ....................................the gas unit.
Temperature ..............................the temperature.
Temperature Unit.......................the temperature unit.
Barometric Pressure..................the barometric pressure.
Barometric Pressure Unit ..........the barometric pressure unit.
Event .........................................the event bit mask in hexadecimal format.
The values are not described here. Refer to
List of events and alarms on page 105
.
•
Example of one measurement:
CH1 697.176 ppb 20.1 °C 0.982 bar C00
2.
If the option "Mode = expert" is chosen, the following message is sent:
with:
Phase shift.................................The fluorescence phase shift in [°].
Partial pressure .........................The partial pressure in [bar].
Time ..........................................The time of the measurement. Format "hh:mm:ss."
Index..........................................This is the index of the last measurement.
This number starts at 0 at power up of the program. The following is an example of one
measurement:
CH1 697.173 ppb 20.1 °C 0.982 bar 03000000 26.045 ° 0.69700
bar -21.409 ° -64.991 ° 2.349 V 2.499 V 25.531 °C 45.000 22:59:42
5923
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t Barometric Pressure\t Barometric Pressure Unit\t\ Event\t\r\n
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t BarometricPressure\t Barometric Pressure Unit\t
Event\t Phase Shift\t °\t Partial Pressure\t bar\t Reference Phase\t °\t
Fluorescent Phase\t °\t Reference Amp\t V\t Fluorescent Amp\t V\t
Tinst\t Tinst Unit\t Offset\t Time\t Index\r\n
Содержание ORBISPHERE K-M1100
Страница 1: ...DOC024 52 93023 ORBISPHERE Model K M1100 Sensor and Model 410 Analyzer USER MANUAL 01 2021 Edition 18 ...
Страница 5: ...4 Table of Contents ...
Страница 19: ...18 Specifications ...
Страница 39: ...38 Installation ...
Страница 45: ...44 User Interface ...
Страница 48: ...47 Section 6 View Menu Figure 33 View menu ...
Страница 51: ...50 View Menu ...
Страница 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu ...
Страница 59: ...58 Measurement Menu ...
Страница 65: ...64 Calibration Menu ...
Страница 66: ...65 Section 9 Inputs Outputs Menu Figure 39 Inputs Outputs menu ...
Страница 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 40 Analog outputs menu ...
Страница 75: ...74 Inputs Outputs Menu ...
Страница 91: ...90 Communication Menu ...
Страница 94: ...93 Section 12 Products Menu Figure 43 Products menu ...
Страница 97: ...96 Global Configuration Menu ...
Страница 98: ...97 Section 14 Services Menu Figure 45 Services menu part 1 ...
Страница 99: ...98 Services Menu Figure 46 Services menu part 2 ...
Страница 107: ...106 Maintenance and Troubleshooting ...
Страница 113: ...112 Accessories and Spare Parts ...