70
Inputs/Outputs Menu
Analog outputs (continued)
Calibration of the analog
output
The calibration of the analog output is aimed at aligning the internally
calculated current to the real current output. This was performed at factory,
but could become necessary again because of electronic tolerances. A
precision amperometer connected at the corresponding analog output
connection point is required. Refer to
.
•
Select the analog output channel to calibrate and press the
Start
button.
•
Measure with the amperometer the current value for point 1. It should be
below 4 mA
•
Edit point 1 and enter the same value as read on the amperometer, then
press the
Continue
button.
•
Measure with the amperometer the current value for point 2. It should be
above 20 mA.
•
Edit point 2 and enter the same value as read on the amperometer,
before pressing the
Continue
button.
Calibration of the selected analog output channel is completed.
Direct test
Test to check the calibration of the analog outputs. A precision amperometer
connected at the analog output connection point is required.
•
Select a value (4, 12, 20 mA available) for each channel and compare
this value (+ 0.02 mA) with what the amperometer shows.
A calibration is required if the value on the amperometer differs from the
current se 0.02 mA).
Note:
It is possible to test one analog output without interfering with the
others. During the test, the other analog outputs will continue to indicate the
measurement.
Characteristics test
This is a test for the correct operation of the peripherals connected to each
analog output, by verifying that the PLC computes the correct value.
The analog output will send the current corresponding to the value entered
in the text boxes.
•
Type in a test value for each analog output, and check for the related
action on the peripheral.
Содержание ORBISPHERE K-M1100
Страница 1: ...DOC024 52 93023 ORBISPHERE Model K M1100 Sensor and Model 410 Analyzer USER MANUAL 01 2021 Edition 18 ...
Страница 5: ...4 Table of Contents ...
Страница 19: ...18 Specifications ...
Страница 39: ...38 Installation ...
Страница 45: ...44 User Interface ...
Страница 48: ...47 Section 6 View Menu Figure 33 View menu ...
Страница 51: ...50 View Menu ...
Страница 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu ...
Страница 59: ...58 Measurement Menu ...
Страница 65: ...64 Calibration Menu ...
Страница 66: ...65 Section 9 Inputs Outputs Menu Figure 39 Inputs Outputs menu ...
Страница 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 40 Analog outputs menu ...
Страница 75: ...74 Inputs Outputs Menu ...
Страница 91: ...90 Communication Menu ...
Страница 94: ...93 Section 12 Products Menu Figure 43 Products menu ...
Страница 97: ...96 Global Configuration Menu ...
Страница 98: ...97 Section 14 Services Menu Figure 45 Services menu part 1 ...
Страница 99: ...98 Services Menu Figure 46 Services menu part 2 ...
Страница 107: ...106 Maintenance and Troubleshooting ...
Страница 113: ...112 Accessories and Spare Parts ...