FAQ
175
F
AQ
•
The accuracy does not match the specification.
•
The display shows “fuse broken” message.
•
The display shows “output shorted” message.
•
The display shows “waiting for 10secs” message.
•
The display shows “OCP” message.
•
The display shows “Remote Sense Error” message.
The accuracy does not match the specification.
Make sure the device is powered On for at least 30 minutes, within
+18°C~+28°C. This is necessary to stabilize the unit to match the
specification.
The display shows “fuse broken” message.
The internal protection fuse is open. Contact your local dealer or
GW Instek.
The display shows “output shorted” message.
Make sure the DUT is correctly connected to the APS-7000 output
terminals and that no shortcut exist. A DUT malfunction may also
be the cause.
Содержание APS-7000 Series
Страница 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 25: ...GETTING STARTED 25 Indicates that the panel lock is active ARB Indicates that the ARB function is active ...
Страница 89: ...OPERATION 89 Exit 4 Press Exit F4 to exit from the MISC Configuration settings Example Settings ...
Страница 111: ...OPERATION 111 Simulate Example ...
Страница 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Страница 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Страница 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Страница 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Страница 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...