APS-7000 Series User Manual
88
Miscellaneous
The Miscellaneous menu contains miscellaneous parameter settings.
T Ipeak, hold → from page 88
Power ON Output → from page 90
Buzzer → from page 91
SCPI Emulation → from page 92
Remote Sense → from page 93 (for APS-7200 and APS-7300 only).
T Ipeak, hold
The T Ipeak, hold function sets the hold time for the peak current
measurement. After the output is turned on, the APS-7000 will
delay starting the peak current measurement by this hold time.
Output on
Ipeak Hold time
I
Begin the peak current
measurement
T
Steps
1.
Press the
Menu
key. The Menu
setting will appear on the display.
Menu
2.
Use the scroll wheel to go to item 4,
MISC
Configuration
and press
Enter
.
3.
Go to the
T Ipeak, hold(msec)
setting using the
scroll wheel and press
Enter
. Set the time and
press
Enter
again to confirm.
T Ipeak
1 ~ 60,000 ms
Содержание APS-7000 Series
Страница 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 25: ...GETTING STARTED 25 Indicates that the panel lock is active ARB Indicates that the ARB function is active ...
Страница 89: ...OPERATION 89 Exit 4 Press Exit F4 to exit from the MISC Configuration settings Example Settings ...
Страница 111: ...OPERATION 111 Simulate Example ...
Страница 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Страница 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Страница 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Страница 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Страница 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...