Firmware Update
Background
The APS-7000 firmware can be upgraded using
the USB A port on the front panel. See your
local distributor or the GW Instek website for
the latest firmware information.
Note
Ensure the DUT is not connected.
Ensure the output is off.
1.
Insert a USB Flash Drive into the USB port on
front panel of the APS-7000.
The USB drive should include the
gw.sbt
file
in a directory name “gw”(USB\gw:).
Steps
1.
Press the
Menu
key. The Menu
setting will appear on the display.
Menu
2.
Use the scroll wheel to go to item 11,
Special
Function
and press
Enter
.
3.
Key in the password when prompted and then
press
Enter
.
The password is “5004”.
4.
Go to Item 1,
Update Main Program
and press
Enter
.
Содержание APS-7000 Series
Страница 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 25: ...GETTING STARTED 25 Indicates that the panel lock is active ARB Indicates that the ARB function is active ...
Страница 89: ...OPERATION 89 Exit 4 Press Exit F4 to exit from the MISC Configuration settings Example Settings ...
Страница 111: ...OPERATION 111 Simulate Example ...
Страница 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Страница 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Страница 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Страница 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Страница 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...