APS-7000 Series User Manual
84
4.
Set the remaining parameters.
Note: these parameters are not visible when
MODE
is set to
OFF
.
Remaining parameters
ACV, T1, T2
Exit
5.
Press
Exit[F4]
to exit from the
Surge/Dip Control settings.
6.
After exiting the menu, the
surge/dip control icon will
appear in the status bar.
Triggering the
Manual
Surge/Dip Site
For the manual mode, the surge/dip site is
determined by a manual trigger.
1.
Configure the nominal voltage
and frequency settings. See the
Basic Operation chapter for
details.
2.
Turn the output on. The nominal
settings above will be output.
Page 78
3.
Press
Shift
to arm the trigger.
Press
TRIG[F4]
to generate the
trigger manually via the front
panel.
Shift
Or,
Alternatively, pulse the Trigger In
pin on the J1 connector high to
generate the trigger.
OR
Page 150
Note
The TRIG soft-key is only available when the surge/dip
control is set to
Manual
.
Содержание APS-7000 Series
Страница 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 25: ...GETTING STARTED 25 Indicates that the panel lock is active ARB Indicates that the ARB function is active ...
Страница 89: ...OPERATION 89 Exit 4 Press Exit F4 to exit from the MISC Configuration settings Example Settings ...
Страница 111: ...OPERATION 111 Simulate Example ...
Страница 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Страница 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Страница 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Страница 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Страница 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...