OPERATION
131
Jump Cnt
1 ~ 255, 0
Note: A setting of 0 will set the
number of jumps to infinite.
8.
Go to the
Branch 1/2
setting and set a step to
branch to.
Branch 1, 2 ON, OFF, 0 ~ 255
9.
Go to the
Term
setting and set the step
termination setting. CONTI will automatically
go to the next step at the end of the step. END
will return to step 0. HOLD will stay at the
current step until you allow the sequence to
continue to the next step.
Term
CONTI, END, HOLD
10.
Go to the
Trig Out
setting and set the output
state of the trigger out pin when the step has
started.
Trig Out
HI, LO
11.
Go to the
ON Phs
setting and set the starting
phase of the step.
ON Phase
ON, OFF
ON Phase
0~ 359º
Resolution
1º
12.
Go to the
OFF Phs
setting and set the end
phase of the step.
OFF Phase ON, OFF
OFF Phase
0~ 359º
Resolution
1º
Содержание APS-7000 Series
Страница 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 25: ...GETTING STARTED 25 Indicates that the panel lock is active ARB Indicates that the ARB function is active ...
Страница 89: ...OPERATION 89 Exit 4 Press Exit F4 to exit from the MISC Configuration settings Example Settings ...
Страница 111: ...OPERATION 111 Simulate Example ...
Страница 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Страница 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Страница 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Страница 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Страница 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...