Note:
a)
The cleaning and disinfection must be performed within 10min
before use.
b)
The disinfectant should be sprayed immediately after the use but no
foaming is allowed.
c)
In addition to 75% alcohol, non-residue disinfectants such as Oxytech
from Germany can be used, but the concentration, temperature and time
specified by the disinfectant manufacture must be followed.
6.4
The cleaning, disinfection and sterilization of contra-angle, lip
hook, file clip, protective silicon cover, touch probe are as follows.
Unless otherwise stated, they will be hereinafter referred to as
“products”.
Warnings:
The use of strong detergent or disinfectant (alkaline pH>9 or acid pH
<5) will reduce the life span of products. The manufacturer takes no
responsibility if the recommendations are not followed.
The products should not be exposed to temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization
cycles. The materials used in manufacturing process were selected
accordingly. However, with every cycle thermal and chemical stresses will
result in ageing of the products. The maximum number of sterilizations
for products is 250 times.
6.4.1
Initial processing
6.4.1.1
Processing principles
Effective sterilization is possible only after thorough cleaning and
disinfection. Please ensure sterility of products during use, only
sufficiently validated equipment and product-specific procedures are used
for cleaning/disinfection and sterilization, and that the validated
parameters are adhered to during every cycle.
Please also adhere to legal requirements applicable in your country as
well as the hygiene regulations of the hospital or clinic, especially with
regard to the additional requirements for the inactivation of prions.
6.4.1.2
Post-operative treatment
The post-operative sterilization must be carried out immediately, no later
than 30 minutes after the completion of the procedure. The steps are as
follows:
Содержание Ai-Motor MotoPex
Страница 1: ...Please read this User Manual carefully before use for operating instructions care and maintenance ...
Страница 4: ...2 ...
Страница 5: ...3 ...
Страница 42: ...40 the contra angle ...
Страница 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Страница 56: ...ZMN SM 034 V1 9 20210430 ...