17
and press Main button to entry handpiece functions setting, press Setting
button “P” till target setting, press Adjusting button “+”/“-” to adjust, then
press the Main button to confirm.
3.2
Screen display
Standby interface
a.
Customized program sequence
number 0-9, total 10 programs.
b.
Battery level
c.
Set speed
d.
Set torque
e.
Operation mode
Working interface
a.
Set speed
b.
Set torque
c.
Real time torque
d.
Torque display scale
Working length measurement mode
interface
a.
Apical reference point flash bar
b.
EAL: Electronic apex locator
Working Length measurement state
interface
a.
Canal length indicator bar
b.
Indication number
Digital numbers 00-16 do not
represent the actual length from the
apical foramen. It simply indicates
the file progression towards the
apex. Number “00” indicate that the
file has reached the apical foramen.
c.
Apical foramen.