23
Only activating in REC and ATR operation
mode.
F: Forward Angle. In the REC mode, the
Forward Angle of 20°~400° is available. In the
ATR mode, the Forward Angle of 60°~400° is
available.
R: Reverse Angle In the REC mode, the
Reverse Angle of 20°~400° is available.
In the ATR mode, the reverse Angle cannot be
greater than the forward Angle.
Press Adjusting button “+”/ “-” to change the
angle,
adjustable every 10 degrees.
It is suggested that the difference between the
forward angle and reverse angle should be
greater than or equal to 120 degrees, root to
prepare root canals effectively.
Forward Angle<Reverse Angle, such as F:
30º/R: 150º, the effective cutting angle is
Reverse Angle, is suitable for the reciprocating
files such as WOODPECKER W3-ONE.
Forward Angle>Reverse Angle, such as F:
180º/ R: 30º, the effective cutting angle is
Forward Angle, is suitable for the reciprocating
files such as SENDONELINE S1.
4.3
Preset program selection
For convenience, some file systems are
preset. Press the Adjusting button “+”/ “-” to
switch to the preset program (M0-M9, preset
program 1-5 ), the interface will be displayed
as depicted on the left side of the row.
press Setting button “P” for a long time to enter
the preset program during standby state, the
interface will be displayed as depicted on the left
side of the row. Press the Adjusting button “+”/
“-” to select the filesystem.
Содержание Ai-Motor MotoPex
Страница 1: ...Please read this User Manual carefully before use for operating instructions care and maintenance ...
Страница 4: ...2 ...
Страница 5: ...3 ...
Страница 42: ...40 the contra angle ...
Страница 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Страница 56: ...ZMN SM 034 V1 9 20210430 ...