39
1.
Remove the products from the base, and rinse away the dirt on the
surface of the handpiece with pure water (or distilled water/deionized
water);
2.
Dry the products with a clean, soft cloth and place them in a clean
tray.
Notes:
a) The water used here must be pure water, distilled water or deionized
water.
6.4.2 Preparation before cleaning Steps:
Tools: tray, soft brush, clean and dry soft cloth.
1.
Remove the files from the contra-angle.
2.
Remove the file clip, isolation sleeve, Contra-angle and
connecting wire from the handpiece in sequence and place them in a
clean tray.
3.
Use a clean soft brush to carefully brush the lip hook, file clip,
protective silicon cover, touch probe, head and back cover of the contra-
angle until debris or stains on the surface are not visible.
4.
Use a soft cloth to dry the products and place them in a clean tray.
The cleaning agent can be pure water, distilled water or deionized water.
Disassembling steps
a)
Press the push-button and pull out the shank/file.
b)
When removing the protective silicon cover, pull it out slowly.
c)
Make sure the handpiece is switched off before Inserting or removing